Gas
Fitting Types
It is absolutely critical that only trained
personnel disconnect and re-connect medical gas cylinders. Mishandling, not
connecting the cylinders correctly, or cross connecting gases can lead to
serious injury and even death.
-
do not attempt to modify the cylinder connections
-
always verify that the gas cylinder label matches the
CGA fitting type for that particular gas
-
only trained personnel should replace cylinders
-
never remove or change the fittings on the gas
cylinders
|
Oxygen (O2):
NFPA 99
color: green
connection type: CGA-540
|
 |

click image to enlarge |
|
Air:
NFPA 99 color: yellow
connection type: CGA-346
|
 |

click image to enlarge |
|
Nitrogen (N2):
NFPA 99 color: black
connection type: CGA-580
|
 |

click image to enlarge |
|
Nitrous Oxide (N2O):
NFPA 99 color: blue
connection type: CGA-326
|
 |

click image to enlarge |
|
Carbon Dioxide (CO2):
NFPA 99
color: gray
connection type: CGA-320
|
 |

click image to enlarge |
|
Medical Breathing Mixtures:
NFPA 99
color: varies (consult your gas supplier)
connection type: CGA-280
|
|

click image to enlarge |
|
|
|
|
|
© 2006, Medical Gas Info. All Rights Reserved |